• HOME
  • ISSUES
  • EXPLORE CONTENT
    • Introduction
    • Articles
    • Book Reviews
    • Case Reports
    • News
    • Sponsored Editorials
  • EVENTS
    • Introduction
    • Event News
    • Conference Reports
    • Submit your Course or Conference
  • AUTHORS
    • Introduction
    • ACNR Author Guidelines
    • Author Disclosure
  • RESOURCES
  • ABOUT
    • Introduction
    • How is ACNR Funded?
    • Editorial Board
    • Peer Reviewers
    • ACNR Policies and Forms
    • Ethics and Malpractice
    • Email Newsletter
    • Journal Signup
    • ACNR Privacy Policy
This website is intended for healthcare professionals
ACNR
  • HOME
  • ISSUES
  • EXPLORE CONTENT
    • Introduction
    • Articles
      • Articles
      • History
      • Letters to the Editor
      • Reviews
      • Personal Perspectives
      • View All
    • Book Reviews
    • Case Reports
    • News
    • Project Neurology Webinar Series
    • Sponsored Editorials
  • EVENTS
    • Introduction
    • Event News
    • Conference Reports
    • Submit your Course or Conference
  • AUTHORS
    • Introduction
    • ACNR Author Guidelines
    • Author Disclosure
  • RESOURCES
  • ABOUT
    • Introduction
    • How is ACNR Funded?
    • Editorial Board
    • Peer Reviewers
    • ACNR Policies and Forms
    • Ethics and Malpractice
    • Email Newsletter
    • ACNR Online Community Signup
    • ACNR Privacy Policy
  • Menu

ACNR ARTICLES

Principles of MRI

Publication Date: 30 Apr 2009

Article written by:
Dr Justin Cross

Author

  • Dr Justin Cross

trained in neuroradiology at Addenbroke’s Hospital, Cambridge UK and at the University of Toronto, Canada. He has a special interest in paediatric neuro radiology and has published articles on the measurement of cerebral tumour volume, carotid imaging and the use of spectroscopy in clinical practice.

Correspondence Address:
Dr Justin Cross, Department of Radiology, Box 218, Addenbrooke’s Hospital, Hills Road, Cambridge CB2 0QQ, UK.

Publication Dates:

Publication Date:
30 Apr 2009

Licence:
Creative Commons Attribution


Article Categories

  • ABN (13)
  • ABNT (20)
  • Alzheimer's Disease (8)
  • Aquired Brain Injury (21)
  • Australia/NewZealand (34)
  • Clinical Review (180)
  • Communication aids (1)
  • COVID-19 (16)
  • Dementia (18)
  • Epilepsy (31)
  • Headache (19)
  • History (53)
  • Huntington's Disease (3)
  • Movement Disorders (20)
  • Multiple Sclerosis (25)
  • Myasthenia Gravis (7)
  • Neurological Literature (15)
  • Neurological Signs (10)
  • Neuroopthalmology (5)
  • Neuropsychiatry (7)
  • Neurosurgery (41)
  • Nutrition (5)
  • Paediatric neurology (15)
  • Pain (9)
  • Parkinson's Disease (43)
  • Personal Perspectives (19)
  • Regeneration (3)
  • Rehabilitation (57)
  • Sleep (18)
  • Special Feature (52)
  • Stroke (28)

Basic MRI physics

Frequency and Phase

The rotation of protons can be described in terms of frequency and phase. In a magnetic field, the protons spin around the axis of the magnet at a given frequency which is proportional to the field strength. This is governed by the Larmor equation: frequency = Larmor constant x magnetic field strength.

In addition to frequency, protons rotating in a magnetic field have a property known as phase. In the resting state, the protons are not in phase (Figure 3a). However, when particular radiofrequency pulses are used, the protons can be brought into phase (Figure 3b). When the radiofrequency is turned off, they dephase again. The ability to manipulate the phase of protons is important for differentiating tissues (see T2 weighted imaging) and for the spatial localisation of signal intensity (see spatial localisation).

Understanding spatial frequencies

MR images are acquired and processed in terms of spatial frequency rather than pixel by pixel. This is necessary to overcome the difficulty in localising the origin of MR signal intensity (see spatial localisation). An image is made up of pixels that have amplitude (degrees of greyness) and position. One row of pixels can be represented by a waveform which plots amplitude on the y-axis and distance (position) on the x-axis (Figure 6). A mathematical operation known as a Fourier transform breaks this wave form down into constituent sine waves of differing amplitudes and frequencies. Fourier mathematics can be used to encode a 2D image entirely in terms of spatial frequencies rather than for each row of pixels in turn. This is difficult to conceptualise, but the spatial frequency representation of an image is referred to as k-space.

Spatial localisation

(i) Slice selection

A magnetic gradient is applied along the axis of the magnet. Since protons resonate at frequencies depending on the applied field strength, slices can be selected by varying the radiofrequency energy used to acquire data (Figure 7).

(ii) In slice localisation

We learnt earlier that MR images are obtained in terms of spatial frequency. This followed a break-through in spatial localisation known as phase encoding. The phase of protons was discussed earlier, as well as the ability to manipulate phase using magnetic gradients and radiofrequency. Phase encoding is a brilliant way of localising signal intensity. We have seen that data is being obtained slice by slice by varying the magnetic field along the length of the patient and using a radiofrequency to which only protons in the slice of interest will resonate (Figure 7). How can we localise signal intensity within the slice? Imagine applying a gradient along the x-axis of the slice which causes phase to vary by 3 x 360 degrees across the region of interest (Figure 8). If there is a uniform distribution of protons in this area, signal intensity will cancel out to zero. However, if there are three blocks of protons evenly spaced out, signal intensity from these protons will summate giving a strong signal. This gradient can be said to be sensitive to spatial frequencies of three per unit length. The gradient can be varied to detect any spatial frequency required.

T1 and T2 relaxation times

T1 and T2 are rate constants governing the return of protons to a resting state following excitation by a combination of applied radiofrequency energy and magnetic field changes.

T1 refers to the recovery of the protons’ magnetic field along the axis of the magnet. When radiofrequency is applied, the protons are deflected away from the magnetic field and when it is turned off, the protons re-align with the field. During this process, electromagnetic radiation is emitted. The rate constant governing the return of longitudinal magnetisation is called T1. T1 is governed by the interaction of protons with large molecules such as cell membranes, lipids and myelin and is sometimes referred to as spin-lattice relaxation. Protons that are heavily bound have a rapid recovery of magnetisation and a short T1. There are a few substances that have a particularly short T1 (see Table 1) so that lesions that are high intensity on T1 can be easily characterised. Protons that are free to move, for example in solution, have a long T1.

T2 governs the loss of phase that occurs when a radiofrequency pulse is switched off. T2 is dominated by interaction with other protons (spin-spin interactions). Protons that are closely packed influence each other more than those in free solution. Thus water (very loosely packed) has a long T2 and tissues with little water content have a short T2. T2 is particularly useful in detecting the water content of tissues and is sensitive to pathologies that generate oedema (inflammation, infection, ischaemia, neoplasia etc).

Artefacts

Ghosting

If errors are made in the detection of spatial frequency data, ghosts of the image are generated usually to the left and right side of the main image in the phase encoding axis. This can occur because of patient motion or from flowing blood which has a different phase to the surrounding tissues. The mathematical explanation for this phenomenon is beyond the scope of this article but can be found in an MR physics text.

Gibbs or truncation artefact

We saw earlier that MRI has difficulty in representing straight lines and sharp edges. When there is a straight line, such as the interface between the cord and CSF, multiple lower intensity lines may be repeated on either side of the interface. This can be mistaken for an intrinsic cord abnormality. The artefact is named after Willard Gibbs (1839-1903) an American physicist who first described this phenomenon in relation to Fourier transforms. It is also referred to a truncation artefact because the problem would not arise if an infinite number of spatial frequencies were used (in practice the number of frequencies is truncated to allow a reasonable imaging time).

| Sign up to the ACNR email newsletter here |

Phase wrap

If the field of view selected is smaller than the slice, signal from outside the field of view may be projected on top of the desired image. This artefact can be avoided by selecting a larger field of view or nulling the signal from areas that are not of interest.

Susceptibility

If there is material in the region of interest which modifies the magnetic field, signal in adjacent tissue is modified and spatially misrepresented. This leads to distortion of the image, usually with loss of signal in adjacent tissues.

Chemical shift

Protons in fat resonate at slightly different frequencies to protons in water when the same magnetic field strength is applied. This results in slight spatial misregistration of signal from fat. 


Figure 1 – MRI equipment
• A superconducting magnet made of millions of turns of wire consisting of Niobium-
Titanium alloy embedded in copper. This is maintained at a temperature below 7.7K
by liquid helium which in turn is kept cold by a refrigeration unit (cold head). Helium
is slowly lost from the system and must be refilled every 2-3 years.
• Transmit-receive coils made of copper which emit radiofrequency energy when an
electrical current is passed through them. When radiofrequency passes through
them an electrical current is induced and this is how magnetic resonance signal
intensity is detected.
• Motorised table
• Faraday cage – copper wires are built into walls and ceiling to prevent extraneous
radiofrequency energy entering the room.
• Ventilation is required to maintain oxygen levels, which can become depleted if heli-
um slowly leaks into the room even during normal operation of the magnet. If cool-
ing of the magnet fails, the coils will become resistive causing a sudden rapid rise in
temperature, a process known as quenching. This will result in a large release of heli-
um gas, and facilities must be in place to allow this to be safely vented to the out-
side of the building.



Figure 4 – T1 weighting
This is sometimes referred to as spin-lattice imaging as it is influenced by the interaction of protons with large molecules. These molecules tend to shorten T1 and this effect is displayed as high intensity on MR images. On this T1 weighted image, subcutaneous fat and myelin are shown as high intensity; grey matter and muscle are intermediate intensity; water is low intensity; bone and air are very low intensity.


Figure 5 – T2 weighting
T2 weighted imaging may be referred to as spin-spin imaging which describes the effect of neighbouring protons on each other. T2 refers to the rate at which the phase of protons is lost over time. In practice, T2 hyperintensity is strongly influenced by water content, although fat is also hyperintense. On this T2 weighted image, CSF and subcutaneous fat are high intensity; grey matter is higher intensity than white matter; bone and air are very low intensity.

Reference

McRobbie DW, Moore EA, Graves MJ, Prince MR. proton. Cambridge University Press, 2003.



Related

  • Published by Whitehouse Publishing
  • The Lynch, Mere, Wiltshire BA12 6DQ, United Kingdom
  • info@acnr.co.uk
  • +44 1747 860168

FOLLOW US

  • Facebook
  • Instagram
  • X
  • Linkedin

RSS FEEDS

  • ACNR RSS
  • All
  • Articles
  • Events

ACNR SIGNUPS

Join our email list to be notified when there is new content or upcoming events.

© 2026 ACNR. Published by Whitehouse Publishing. Read our Privacy and Cookie Policy. Website by Sugar Web.

SIGN UP FOR OUR EMAILING LIST

Sign up to receive our email newsletter with links to the latest content. ACNR is free, thanks to the support of advertisers. The editorial content is peer reviewed and remains completely independent unless clearly specified.

For UK-based specialists, sign up for our journal mailing list.

WARNING

URL here

You are now leaving the ACNR website, please click the "Continue" button below. If you do not wish to leave, please close this pop-up.

Continue